RESEARCH ARTICLE


LEARNING FROM IMAGES – Dysfunctional Dialysis Catheter



Tushar J. Vachharajani*


Article Metrics

CrossRef Citations:
0
Total Statistics:

Full-Text HTML Views: 2355
Abstract HTML Views: 1942
PDF Downloads: 808
ePub Downloads: 974
Total Views/Downloads: 6079
Unique Statistics:

Full-Text HTML Views: 1182
Abstract HTML Views: 1026
PDF Downloads: 491
ePub Downloads: 480
Total Views/Downloads: 3179



Creative Commons License
© Tushar J. Vachharajani; Licensee Bentham Open.

open-access license: This is an open access article licensed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted, non-commercial use, distribution and reproduction in any medium, provided the work is properly cited.

* Address correspondence to this author at the W.G. (Bill) Hefner VAMC, 1601 Brenner Ave, Salisbury, NC 28144, Tel: 704-638-9000; Fax: 704-638-3855; E-mail: tushar.vachharajani@va.gov




Tunneled dialysis catheter is a frequently used vascular access to initiate hemodialysis treatment for patients with end stage renal disease. Right internal jugular vein is the preferred site due to its straight continuation in to the superior vena cava [1]. A well positioned tunneled catheter will have a smooth transition from the tunnel in to the internal jugular vein. Improper placement technique can lead to early dysfunction and inadequate dialysis treatment. Fig. (1A) shows an acute kink in the tunneled portion of the catheter resulting in catheter dysfunction. The ideal treatment option, when possible, is to create a new tunnel utilizing the same venous puncture thus preserving the venous access site. Additionally, the procedure is short with minimal discomfort to the patient. Fig. (1B) shows a new catheter with a smooth curve at the transition point from the subcutaneous tunnel into the internal jugular vein.

Fig. (1).

(A) A right internal jugular vein tunneled dialysis catheter with a kink (marked by arrow). (B) Well positioned tunneled dialysis catheter utilizing the same venous puncture site. Old catheter has been cut at the venous puncture site (marked by chevron).


CONFLICT OF INTEREST

The author confirms that this article content has no conflict of interest.

ACKNOWLEDGEMENT

Declared none.

REFERENCES

[1] Vachharajani TJ. Atlas of Dialysis Vascular Access Available at: www.fistulafirst.org 2012 [Accessed: July 22, 2012];